PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
74
tolerance smaller reduces the time required to float the piston and making the tolerance
wider increases the time. However, if the tolerance is set too small so that it does not
correctly reflect the degree of agreement between the pressure controller pressure
measurement and the PG9000’s floating piston, overshoot of the piston float point or
inability to float the piston will result.
The pressure controller tolerance determines, when applicable:
a) the distance from the actual pressure target that pressure is set before the
AutoGen function begins to seek movement of the piston.
b) the amount of pressure overshoot that must occur before “forced rotation” is
used to overcome possible friction between the piston and the cylinder.
OPERATION
To view or set the pressure control tolerance, press
[GEN]
,
<5tol>
.
The current value of the pressure control tolerance is displayed. The value is in
% of full scale of the pressure controller. If the pressure controller is
AutoRanged, the AutoRange maximum pressure is used as full scale (see the
pressure controller’s Operation and Maintenance Manual).
The default tolerance value is 0.05% of full scale.
Edit the tolerance value if desired. Do not make the pressure tolerance less than
the worst case agreement between the pressure controller pressure measurement
and the pressure indicated by the PG9000 when its piston is floating.
3.10.9.5 <6refloat>
PURPOSE
To turn ON and OFF a function that causes the automated generation component (see
Section 3.10.9) to refloat the piston to the target piston position after a new pressure or
mass target is entered, even if the piston is already floating within the piston position
Ready
limits. When Refloat is ON, the piston is always refloated to the target piston
position after a new pressure or mass target command. This gives the full stroke of the
piston to drop before refloat is necessary. When Refloat is OFF, the time required to
activate the pressure control component and refloat the piston is eliminated if it is not
needed. This can result in very rapid pressure setting when the piston is still floating
after a mass load change.
OPERATION
To turn ON and OFF the Refloat function, press
[GEN]
,
<6refloat>
. The cursor is on
the choice corresponding to the current state. Select
<1no>
for the piston NOT to be
controlled to the target position after a pressure or mass command if the piston is
already in the
Ready
piston position limits. Select
<2yes>
for the piston to always to be
controlled to the raise the piston. The default is
<2yes>
.
3.10.10 [RES]
PURPOSE
To set the mass loading resolution with which PG9000 loads mass in response to pressure or
mass commands (see Section 3.10.12). The resolution is commonly set to 0.1 kg. This does
not affect the pressure display resolution as it will always be displayed to the highest
resolution of digits.
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...