![Fluke PG9000 Series Operation And Maintenance Manual Download Page 12](http://html1.mh-extra.com/html/fluke/pg9000-series/pg9000-series_operation-and-maintenance-manual_2309961012.webp)
PG9602™ OPERATION AND MAINTENANCE MANUAL
Page 2
© 2011 Fluke Calibration
1.2
Specifications
1.2.1
General Specifications
Power Requirements
50/60 Hz, 60 VA max. consumption.
100 to 240 VAC, operating altitude: <2000m
100 to 150 VAC, operating altitude: <5000m
Operating Temperature Range
15 to 35 °C
Operating Humidity Range
5 to 95% R.H., non-condensing
Weight
Instrument platform with no mass or
piston-cylinder loaded.
PG9602 (without vacuum hardware)
Remote Electronics Module
PG Terminal
Optional Vacuum Reference Hardware
34 kg (74 lb)
1.9 kg (4 lb)
1.4 kg (3 lb)
16 kg (36 lb)
Dimensions
PG9602 Base (without vacuum hardware)
PG9602 Base (with optional vacuum bell
jar and vacuum gauge connected)
Remote Electronics Module
PG Terminal
34.1 cm H x 54.3 cm W x 52.3 cm D (13.4 in x 21.4 in x 20.6 in)
(Height: top of piston-cylinder assembly)
56 cm H x 54.3 cm W x 52.3 cm D (22 in x 21.4 in x 20.6 in)
(Height: Top of bell jar)
10.2 cm H x 35.1 cm W x 19.6 cm D (4 in. 13.8 in. x 7.7 in.)
12 cm H x 15 cm W x 20 cm D (4.7 in. H x 5.9 in. W x 7.9 in. D)
Microprocessors
Instrument Platform
PG Terminal
Motorola 68302
Hitachi 64180
Communication Ports
RS232
IEEE-488
COM1: Host computer
COM2: Residual vacuum sensor (external)
COM3: Automated pressure generator/controller
COM4: Unused/spare or pass through
Host computer
Fuses
PEM Fuse: 1A, 250V, 5x20mm, Slow blow
Internal power supply Fuse: 3.15A, 250V
Overall Pressure Range
Gauge:
12 kPa to 11 000 kPa (1.7 to 1 600 psi)
Absolute:
12 kPa to 11 000 kPa (1.7 to 1 600 psi)
Differential:
- 95 to 985 kPa (-13.7 to 143 psi) at
15 to 110 kPaa (2.2 to 16 psia) static pressure
Operating Media
Gas: nitrogen, helium, dry air (dew point
≤
-40 °C)
Maximum Mass Load
100 kg
1
, while not exceeding 11 MPa (1 600 psi)
2
1
MS-AMH-100 mass set contains approximately 104.4 kg of mass.
Choice of piston-cylinder will add 0.2 to 0.5 kg of mass.
2
The PC-7100/7600-200 piston-cylinder is limited to use of 55 kg.
Pressure Connections
PG 9602 Base
AMH Drive
AMH Vent
Remote Electronics Module
Vacuum Connections (Bell Jar)
Test port: DH200
Quick connector equivalent to Swagelok QM Series (QM2-B-200),
Use with DESO (double end shut off) type stem
Quick connector equivalent to Swagelok QM Series (QM2-B-200),
Use with SESO (single end shut off) type stem only.
Note
DH200 is a gland and collar type fitting for 1/4 in. (6.35
mm) coned and left hand threaded tubes.
DH200 is equivalent to AE SF250C, HIP LF4, etc.
ATM port:
10-32 UNF
3 each KF40, 1 each KF50 (top)
CE Conformance
All PG9000 models conform to CE requirements and are CE
marked.
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...