PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
114
3.12.8.1 <2control>, <1up/down>
PURPOSE
To move the AMH mass load to the up or down position (see Section 3.7.1).
OPERATION
To move the mass handler actuator to the up and down positions, press
[SPECIAL]
,
<8AMH>
,
<1up/down>
.
The display indicates the current position. Press
[ENT]
to change the position.
This control does not actually change the state of the binary mass pins nor the
main mass column positions.
Warning
Do not put fingers or anything else under the AMH trim mass
tray. When the mass load is lowered, it can pinch or crush
anything under it.
3.12.8.2 <2control>, <2discreet>
PURPOSE
To load and unload specific AMH masses.
OPERATION
To load and unload specific AMH masses, press
[SPECIAL]
,
<8AMH>
,
<2control>
,
<2discreet>
.
The display is:
1* 2 4* 8 16 32
64 Main:1
The
<1>
,
<2>
,
<4>
,
<8>
,
<16>
,
<32>
and
<64>
indicate the 0.1, 0.2, 0.4, 0.8, 1.6,
3.2 and 6.4 kg binary tube masses.
<*>
indicates that the mass is loaded. Move
the cursor to a mass and press
[+/-]
or
[.]
to change its status.
<Main:>
indicates the number of main masses that are loaded. Edit this number
to the desired value.
If any condition in the discreet display is changed the bottom right of the display
changes to
<ENT loads>
. Pressing the
[ENT]
key
to causes AMH to operate to
load the new mass configuration. Press
[ESC]
to return to the AMH menu with
no change in mass load.
Warning
Do not put fingers or anything else under the AMH trim mass
tray. When the mass load is lowered, it can pinch or crush
anything under it.
3.12.8.3 <2control>, <3loadall>
PURPOSE
To clear the AMH and cause all the masses in the AMH mass set to be loaded onto
the piston. The function is used to remove all the masses from the mass handler
and prepare it to be removed from or installed onto the PG9000 platform.
OPERATION
To load all AMH masses, press
[SPECIAL]
,
<8AMH>
,
<3loadall>
. Confirm the
load all instruction and AMH proceeds to lower all the masses onto the piston.
Warning
Do not put fingers or anything else under the AMH trim mass
tray. When the mass load is lowered, it can pinch or crush
anything under it.
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...