SFC
Functions
4.
87
4.7
Control parameters
Control parameters are input variables that determine the control
functions of a controller. Different parameters apply to each type
of controller.
Please note
The control parameters are listed alphabetically.
Flow rate source
This parameter is only available during single feed forward in
application 2.
This parameter switches off the flow input (off) and activates the
flow rate signal for the single feed forward (factory setting = flow
measurement) as control variable.
For quantity-proportional dosing, the parameter must be set to
„Flow Measurement“.
Flow rate direction
This parameter determines the direction of the flow rate signal
directly proportional to the actuator output:
Max. pulses/min
Meaning:
Maximum number of pulses
Explanation:
The pulses max./min parameter only applies to pulse pumps.
This parameter is used to set the maximum number of pulses per
minute in accordance with the employed pump.
Setting range:
The parameter max. pulses/min. can be set to 100, 120, 140, 160
or 180 pulses.
Max.lin.Corr
This parameter monitors changes to already learned dosing
factors.
If new dosing factor changes are learned, which are larger than the
max. linearity correction, this dosing factor is used for all values in
the dosing curve = > initialization of the curve.
direct
=
Flow rate input signal directly proportional to
the positioner output (factory setting)
inverse
=
1 flow rate input signal
Example:
0 – 100 % flow rate = 0 – 20 mA (direct)
0 – 100 % flow rate = 20 – 0 mA (inverse)
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...