Trouble-shooting and
debugging when the measuring
signal is too low or irregular
1
Remove air bubbles on the membrane (G) by lifting the
membrane sensor; air bubbles prevent the ozone from
diffusing through the membrane (G) and distort the
measurement!
Please note
Air bubbles on the electrode shaft (B) and the membrane cap (E)
are normal following the initial startup and subsequent startups
and they disappear by themselves after one or two days.
2
Replenish electrolyte. Open membrane sensor. To do this,
push the elastomer seal (H) to the side so that air can flow in
through the valve opening (F), and only then screw off the
membrane cap (E). Pour out the electrolyte. Wash the
electrode finger (D) and the membrane cap (E) with clean
(distilled) water and dry with a clean paper towel.
Please note
For further procedures, see “Commissioning“. If the membrane
sensor's measuring signal is still too low or irregular, a new
membrane cap (E) must be used. The membrane sensor then
requires a run-in time of approx. one hour, before a calibration can
be carried out.
4.2.7
mA/V input module
The mA/V input module is used for connecting sensors or external
measurements with mA or voltage output signal. 0/4 – 20 mA
signal or 0 – 10 V input voltage are possible.
Various controller functions are available depending on the
application selected. If the mA/V module is used with the „Process
Control“ option (PC), compound loop and single feed forward are
also available.
As measured value display, the unit and display format can be
freely selected in the menu.
4.2.8
Temperature measurement
The A&C board of SFC has a temperature measurement for
connecting a PT1000 sensor (multi-sensor). This temperature
measurement is used for temperature compensation of the „DES“
module and pH measurement. The temperature is shown on the
main display and can be calibrated if necessary. The measuring
range is 0 – 50°C. The unit may be adjusted to °F.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...