SFC
Operation
6.
205
6.4.3
Membrane sensors calibration
Calibration is nearly identical for all membrane sensors. The
difference lies in the fact that some of the chemicals are measured
manually with a photometer and others with a colour meter.
A 1-point calibration is available to calibrate the total chlorine
measurement and the combined chlorine measurement.
Either the total chlorine or the combined chlorine must be
calibrated.
For the selective ozone, chlorine dioxide and free chlorine
measurements, 1-point or 2-point can be selected in the
calibration menu of the „Calibration mode“.
2-point calibration provides the option to compensate for possible
measuring cell zero point offsets.
Total chlorine
In the calibration of the total chlorine measurement, the free
chlorine value is also essential. For this reason, prior to calibration,
the current free chlorine value must be entered in the menu
calibration-Cl-total "Enter free Cl2". Only then shall the total
chlorine calibration be carried out as follows:
1
Using the soft key "calib", select the menu item "Cl-total" and
press "Enter" to confirm.
2
Select the parameter "calib. total" and press "Enter" to confirm.
3
Open the input menu by pressing "Enter". The actual
measured sensor value is thereby cached and can be used for
later calibration. This is essential for fluctuating measurement
values.
4
Immediately after this extract specimen of sample water and
determine the total chlorine content using a comparative
device.
5
Use the arrow keys to enter the determined value.
6
Save the value by pressing "ENTER".
Combined chlorine (optional)
The calibration of the combined chlorine measurement also takes
into account the free chlorine value. It must therefore be ensured
that the free chlorine value (from external measurement system) is
correctly calibrated. Calibration can then take place as follows:
1
Using the soft key "calib", select the menu item "Cl-comb" and
press "Enter" to confirm.
2
Select the parameters "calib. comb" or "calib total" and press
"Enter" to confirm. It does not matter which value is entered.
3
Open the input menu by pressing "Enter". The actual
measured sensor value is thereby cached and can be used for
later calibration. This is essential with fluctuating
measurement values.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...