52
WT.050.590.000.DE.IM.0719
Functions
SFC
4.
Installing the mV sensor
in the flow block assembly
1
Install the mV sensor in the corresponding opening on the flow
block assembly in theY-style flow through assembly.
2
Screw the sensor cable marked "mV" into place.
3
The mV sensor can be calibrated for the first time after approx.
two to three hours running-in time.
Please note
The calibration must be checked after one day.
4.2.4
Fluoride Measurement
The fluoride measurement with fluoride ionic-sensitive electrodes
is used to continuously determine the fluorides in solutions. The
measurement of the fluoride ions concentration is selective, i. e.
the remaining electrodes do not disturb even at a high surplus. The
measurement medium's pH value must be between pH 4 and pH
8.5, in order to obtain a correct result.
Picture 3
Sensor measuring module
Caution!
Quick and repetitive changes in temperature cause the potential to
change continuously, which can lead to electrode malfunction.
Please note
The reference system is filled with electrolyte. The integrated
storage tank must be replenished routinely with the solution.
The fluoride measuring system consists of:
• Fluoride module with a solidly soldered connecting cable
• Fluoride single-rod electrode
The fluoride module must be installed in the module slot of the
SFC. Various controller functions are available depending on the
application selected.
The electrode can be installed in the flow block assemblies
DEPOLOX
®
5/DEPOLOX
®
Pool/VariaSens or in a separate Y-
style flow through assembly.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...