134
WT.050.590.000.DE.IM.0719
Installation
SFC
5.
5.3.2
Pour in the cell sand (only with DEPOLOX
®
5 and
DEPOLOX
®
Pool)
Caution!
Before opening the cover on the pressurized version always first
release the pressure in the cell body with the drain screw.
1
Close the ball valve on the sample water inlet and outlet
(pressurized version).
2
On the non-pressurized version, remove the protection plugs
on the cell body cover of the 3-electrode cells.
On the pressurized version, unscrew the protection plugs on
the cell body cover of the 3-electrode cells.
3
Fill half a cap from the plastic bottle with cell sand and pour it
into the cell body (approx. 1/2 cm³ cell sand).
4
On the non-pressurized version, replace the protection plugs
on the cell body cover of the 3-electrode cells.
On the pressurized version, screw the protection plugs on the
cell body cover of the 3-electrode cells.
Please note
Make sure that the opening (especially the threads in the
pressurized version) is clean; rinse with water, if necessary.
5
Reopen the ball valve on the sample water inlet and outlet
(pressurized version).
Please note
The system must be recalibrated approx. 4 hours after each time
the cell sand is replaced.
The calibration must be checked after one day.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...