224
WT.050.590.000.DE.IM.0719
Maintenance
SFC
7.
Replacing electrolyte, reference
electrode and diaphragms
(DEPOLOX
®
5)
1
Remove cover of flow block assembly DEPOLOX
®
5.
2
Close the ball valve on the sample water inlet and on the outlet
(pressurized version).
3
Open the drain valve and drain the cell body (hold container
underneath).
4
Close the drain valve when the cell body is empty.
5
Remove the sensors. Loosen the cable union (hold the cable
while doing this as it may not be allowed to rotate).
6
Loosen the lower cap on the 3 electrode cell.
7
Remove the signal cable.
Replacing electrolyte
(DEPOLOX
®
5)
8
Unscrew the upper knurled nut on the electrolyte container.
9
Remove the electrolyte tank out of the cell body from below
using the electrode mount.
Please note
Wash the cell sand out of the electrode mount “Changing sell sand
with the 3-electrode cell DEPOLOX® 5, DEPOLOX
10
Remove the electrolyte container from the electrode mount. To
do this, unscrew the knurled nut in the electrode mount.
11
Pull the electrolyte container upwards out of the electrode
mount.
12
Remove the drain plug from the electrolyte container.
13
Turn the electrolyte container upside down and drainthe KCI
electrolytes by lightly shaking it.
Replacing reference electrode
(DEPOLOX
®
5)
14
If necessary, replace the reference electrode.
The reference electrode can be unscrewed from the
electrolyte container. Lightly wet the O-ring before installinga
new reference electrode.
Replacing diaphragms
(DEPOLOX
®
5)
Please note
The diaphragms, which form the contact between the reference
electrodes and sample water, cannot be cleaned. If the water
quality is very good, the diaphragms can remain installed for up to
three years; they should be replaced thereafter (no exceptions). If
the sample water quality is poor, the diaphragms may be soiled.
This influences the measuring accuracy.
15
Remove both diaphragms from the electrolyte container using
a suitable tool (e.g. tweezers).
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...