204
WT.050.590.000.DE.IM.0719
Operation
SFC
6.
Measurement value calibration
(DPD)
10
After zero point calibration, wait at least 2 minutes.
11
Select the parameter "Cal. Span" and confirm with "Enter".
12
Open the input menu by pressing "Enter". The actual
measured sensor value is thereby cached and can be used for
later calibration. This is absolutely essential for fluctuating
measurement values.
13
Immediately after this extract specimen of sample water.
14
Determine the free chlorine, ozone, chlorine dioxide or
potassium permanganate content, for example, with a
photometer.
15
Use the arrow keys to enter the determined value. Finally save
by pressing "Enter". Should the measured sensor value (µA)
according to step 12 have changed, the measured value
displays another value immediately after calibration than that
which was entered in the calibration menu. This is based on
the change in the measured value during the calibration
process between step 12 and 15.
16
Press the soft key "back" several times until the main display is
shown. This concludes the calibration.
Please note
In the DPD calibration of the Cl
2
++ measurement the calibration
value should be greater or equal to 25 % of the measuring range.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...