SFC
Functions
4.
59
5
Rinse off the escaped electrolyte with water.
Please note
The membrane cap (E) must be completely screwed (hand-tight)
onto the electrode shaft (B), so that no gap remains between the
two! After a run-in period of about one hour, the membrane sensor
is sufficiently run-in for an initial calibration to take place.
Calibration should be repeated after one day.
Inserting into the flow-through
adapter
1
Insert the membrane sensor through the cover into the flow-
through adapter until it is resting mechanically on the inflow
mating connector and therefore receives a good flow. It may
be necessary to turn the flow-through adapter cover to set it in
the right position relative to the inflow mating connector.
Please note
Remove air bubbles from the membrane by lifting the membrane
sensor. They interfere with the measurement!
2
Connect the measuring signal cable to the measuring device.
Please note
If the measuring sensor is installed this way, it will function reliably
for approx. three to six months.
Storing the membrane sensor
1
Lift elastomer seal (H) and only then screw off the membrane
cap (E).
2
Wash the membrane cap (E) and electrode finger (D) with
clean (distilled) water.
3
Carefully dry the electrode finger (D) with absorbent paper.
4
Leave the membrane cap (E) to dry in a dust-free place.
5
Screw the dry membrane cap (E) loosely onto the electrode
shaft (B).
Please note
The membrane (G) may not touch the gold working electrode (J).
Restarting
See „Commissioning“.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...