228
WT.050.590.000.DE.IM.0719
Maintenance
SFC
7.
7.5
Maintaining pH electrode
Clean and calibrate if there are fluctuations in the measured
values. A routine schedule cannot be given here because the
cleaning schedule depends heavily on the general condition of the
sample water. In general, calibrate approx. every 4 weeks.
Remove dirt on the glass membrane and diaphragm to prevent
measuring errors.
In particular, the ceramic electrode diaphragm may also be soiled
or coated with lime deposits.
Remove contaminants deposited on the surface of the membrane
glass; use diluted hydrochloric acid (up to 10 %), if necessary.
The electrodes should not be cleaned in a dry state because this
is more likely to smear the layer of dirt over the surface rather than
removing it. Under no circumstances may the membrane be
treated with abrasive cleaning agents.
The electrode must be rinsed subsequently with nothing other than
water.
Remove lime deposits on the glass membrane and the diaphragm
by immersing the electrode into hydrochloric acid (up to 10 %).
Rinse thoroughly with water or distilled water here as well; pH
electrodes age. This is often the cause for a slow display of the pH
value or a drop in the span. pH electrodes typically last 1 to 2
years. However, routine maintenance of the electrodes
recommended.
7.6
Maintaining fluoride electrode
Routinely check the electrolyte level in the electrode (at least once
per week). The fill level should always be just under the filling hole,
approx. 25 mm above the sample water. Replenish the electrolyte,
if necessary.
Routinely calibrate the measuring system to guarantee safe
operation and accuracy.
The electrode shaft can be cleaned using a moist cloth.
Do not touch the glass surfaces with the cloth.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...