SFC
Operation
6.
209
6.4.5
Redox calibration (mV)
Please note
Redox sensors have long running-in times. After calibration with a
buffer solution, it can therefore take several hours until the value
has stabilized.
1
Starting from the basic display in the main menu select the
"Calibration" menu.
2
Select menu item "ORP" and conform with the Enter key.
3
Confirm the menu „Calibrate mV“ with the „Enter“ key.
4
Place one of the supplied beakers into the bottom clip and fill
with the buffer solution or clamp a bag with buffer solution into
the bottom clip.
5
Pull or unscrew the Redox sensor from the lid of the cell body.
6
Dip the Redox sensor through the top clip at least 2 cm deep
into the buffer solution until the indicated pH value remains
constant.
7
Confirm the selection with „Enter“ and enter the ascertained
value with the arrow keys.
8
Save the value by pressing the „Enter“ key.
9
Remove the Redox sensor from the top clip.
10
Insert or screw the Redox sensor into the lid of the cell body.
Redox calibration is then concluded.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...