SFC
Operation
6.
175
Select the µA signal measuring range for 3-electrode cells
and membrane sensors:
70 µA, 100 µA, 200 µA, 1000 µA
for Micro 2000 and Deox 2000: 10 µA, 100 µA, 1000 µA
Allocation of the signal used for reading the free Cl
2
value, either
CAN external or analog (mA input 1).
Only applicable is using a CI-N-sensor.
Setting the connected measurement signal:
0 – 20 mA, 4 – 20 mA, 0 – 10 V, CAN external
power supply should be arranged in accordance with this setting
(see 9. “Wiring Diagrams”)
Factor for adapting an external input signal:
0.1 – 4.0
Adjustment of the reference temperature for the conductivity
measurement:
20°C / 25°C
Display
=> NaCl in g/l / NaCl in % / TDS in g/l / off
TDS factor
=>0.4 – 1.0 (if TDS display is active)
TDS:
Total Dissolved Solids = filtrate dried solid matter content in g/l
The TDS factor depends on the composition of the sample water
and must be determined for each water type.
µA measuring range
Cl2
Mem
Signal Cl
2
free
Mem
Signal
mA/V
Factor
mA/V
Reference temp.
LF
Salt displ.
LF
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...