SFC
Operation
6.
215
Setpoint?
Due to modification of the
measuring range the setpoint of the
controller
is outside the measuring range.
Reset the controller setpoint or adjust
the measuring range
Sensor temp?
Error in the temperature
measurement of the sensor
measurement module/SiDiSens
Interruption in the temperature
sensor or cable
Check the temperature sensor and
cable of the measurement module
Cl
2
comb?
Free Cl
2
measurement via CAN bus
is not available
Check SiDiSens DES, check setup of
CAN bus
Cl
2
++?
pH<6 or pH>8,75
pH measurement via CAN bus is not
available
Check pH measurement, check
SiDiSens DES, check setup of CAN
bus
Cell ?
In 3-electrode cells
Chlorine sensor not screwed in.
No grit cleaning.
Sensor, sensor cable or sensor
module defective.
Sensor measuring module µA
measuring range exceeded
With pH, F
-
and mV modules
Sensor, sensor cable or sensor
module defective
In membrane sensors
Sensor, sensor cable or sensor
module defective
Sensor measuring module µA
measuring range exceeded
In conductivity modules
Sensor, sensor cable or sensor
module defective
Screw in sensor correctly.
Check grit cleaning.
Check sensor, sensor cable or sensor
module, replace if necessary.
Select higher µA measuring range
Check the sensor, sensor cable and
sensor module, replace if necessary
Check the sensor, sensor cable and
sensor module, replace if necessary
Check the sensor, sensor cable or
sensor module, replace if necessary;
clean sensor
Error message
Cause
Remedy
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...