SFC
Operation
6.
203
6.4.2
3-electrode cell (DEPOLOX
®
5 and DEPOLOX
®
Pool)
calibration
Calibration of the 3-electrode cell for Cl
2
, KMnO
4
, O
3
, ClO
2
and
Cl
2
++ (pH-compensated).
Before Cl
2
++ calibration, it must be ensured that the pH
measurement is calibrated correctly.
During calibration of the 3-electrode cell, perform a zero point
calibration and a measurement value calibration (DPD).
The calibration process is nearly the same for chlorine, chlorine
dioxide, ozone and potassium permanganate.
The difference lies in the fact that some of the reagents are
measured with a photometer and others with a colour meter.
Zero point calibration
1
Starting from the basic display in the main menu select the
"Calibration" menu.
2
Select the measurement to be calibrated from the menu, e.g.
Cl
2
free.
3
Select the "Cal. zero" parameter.
4
Close the ball valve on the sample water inlet.
Please note
When the sample water supply has been stopped, the display first
drops rapidly, and after approximately one minute slowly
approaches zero. During commissioning it is essential to wait for 5
minutes, even if the display is already blinking or stands at "0.00"
after a few seconds.
5
Wait until the displayed value or the µA-sensor signal no
longer changes for at least one minute.
6
Press the "ENTER" softkey to set the display to zero and
press „Enter“ again to save the value.
7
Press the softkey „Back“.
Zero point calibration with
disinfectant-free water
8
If disinfectant-free water (e.g. by switching off the dosing
system) is available, zero point calibration can be performed
with it. To do this, switch off the dosing system and perform
steps 1, 2, 3, 5 and 6. The delay times for sample water
extraction and dosing must be observed here (waiting time)!
9
Open the ball valve on the sample water inlet.
Summary of Contents for WALLACE & TIERNAN SFC
Page 1: ...WALLACE TIERNAN MEASURING AND CONTROL SYSTEM SFC from version V 3 08 INSTRUCTION MANUAL...
Page 2: ...2 WT 050 590 000 DE IM 0719 SFC Please note Original instruction manual...
Page 40: ...40 WT 050 590 000 DE IM 0719 Description SFC 3...
Page 118: ...118 WT 050 590 000 DE IM 0719 Functions SFC 4...
Page 121: ...SFC Installation 5 121 Top hat rail assembly...
Page 122: ...122 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 123: ...SFC Installation 5 123 Top hat rail assembly...
Page 124: ...124 WT 050 590 000 DE IM 0719 Installation SFC 5 Wall mounting assembly...
Page 125: ...SFC Installation 5 125 Control cabinet installation assembly...
Page 126: ...126 WT 050 590 000 DE IM 0719 Installation SFC 5 Control cabinet installation assembly...
Page 127: ...SFC Installation 5 127 Installation of the front panel on the control cabinet...
Page 156: ...156 WT 050 590 000 DE IM 0719 Installation SFC 5...
Page 218: ...218 WT 050 590 000 DE IM 0719 Operation SFC 6...
Page 232: ...232 WT 050 590 000 DE IM 0719 Maintenance SFC 7...
Page 266: ...266 WT 050 590 000 DE IM 0719 Complete Units Retrofit Kits and Spare Parts SFC 8...
Page 274: ...274 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 275: ...SFC Declarations and certificates 10 275 10 2 Certificate of Compliance...
Page 276: ...276 WT 050 590 000 DE IM 0719 Declarations and certificates SFC 10...
Page 284: ...284 WT 050 590 000 DE IM 0719 Settings Table SFC 11...