6 Start-up
Hauser
75
6.5.17
Diagnosis
To enter the menu, proceed as follows:
⇒
⇒
Diag
Edit
Next
Select
E
Hold
Error list
Error log
Operating log
Calibration log
Service
pH 7.00
CODE
CHOICE
(Default = bold)
INFO
U
Error list
Error log
Operation log
Calibration log
Service
Error list
: Displays the current active errors.
(Complete error list with description s. page 93)
Error log
: Lists the last 30 signalled errors with date and
time.
Operation log (service code necessary)
: Lists the last
30 registered operating steps with date and time.
Calibration log
: Lists the last 30 calibrations undertaken
with date and time.
!
Note!
• Use the arrow keys to scroll through the lists.
• Leave the lists by pressing "E".
Service:
Y
Factory settings
Simulations
Check systems
Reset
DAT handling
Instrument version
Factory function
Select service diagnosis
Factory settings
: Different data groups can be reset to
the factory settings.
Simulations
: The transmitter behaviour can be simula-
ted after entering of different parameters.
Check systems
: The instrument functions (display,
keys, etc.) can be tested individually.
Reset
: Device reset (no data loss!)
DAT handling
: Copy data into/out of the DAT module.
Instrument version
: Device-internal data e.g. serial
number can be queried.
Factory function
: reset counter, write access
Factory settings:
YA1
Cancel
Settling data
Calibration data
All data
Service data
Operation logbook
Error logbook
Calibration logbook
Set default
Here you can select the data which you wish to
reset to the factory settings.
!
Note!
Danger of data loss. Selecting a point and confirming
with "Enter" deletes all the settings you made in this
area!
Pressing
Cancel
leaves this field without changing the
values.
Calibration data: All the saved data for calibrations
such as zero point, slope, and offset.
Setting data: the remaining data to be set.
All data: calibration data + setting data
Service data: all data + lo reset counters.
Service data / logbooks:
these functions are only for authorised service