![Endress+Hauser Mycom S CPM 153 Operating Instructions Manual Download Page 82](http://html1.mh-extra.com/html/endress-hauser/mycom-s-cpm-153/mycom-s-cpm-153_operating-instructions-manual_2408552082.webp)
6 Start-up
Mycom S CPM 153
82
Hauser
Calibration
redox absolute
"Absolute data entry"
The transmitter has a calibrated mV display range.
One
absolute mV value is set with
a single buffer solution (adaptation of the measuring chain offset). A buffer solution
preferably with 225 or 475 mV is used.
CODE
CHOICE
(default = bold)
INFO
C1
Electrode 1
Electrode 2
shared
Abort calibration
Selection for calibration
(only two-circuit)
Select electrode 1
or
2, and then run through
calibration for each individual electrode.
CB
Calibration with
abs. data entry
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
CBA1 / CBA2
0000 mV
(–1500 ... +1500 mV)
Entry of offset value circuit 1 / 2
Enter the mV value for the electrode offset (electrode
offset = deviation of the measured value display from
buffer solution mV value)
Confirm by pressing
F
. The entered value is effective
immediately.
The maximum offset is 400 mV.
CBA3
Offset too high / too low
Note in display:
Error message if the entered offset leaves the
maximum range.
CBA4
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing
F
accepts the new calibration data.
Reject
: The data is not accepted and the device has
not been recalibrated.
Recalibrate
: The data is rejected and the device is
recalibrated.
CBA5
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that
measurement can take place?