Mycom S CPM 153
8 Troubleshooting
Hauser
95
E027
Failure
Compressed air fail-
ure
Pressure below permitted minimum
yes
no
no
E 030
Failure
SCS fault reference
electrode 1
Reference impedance too high:
Check reference element and, if neces-
sary replace reference or combination
electrode
For IsFET sensor: leak current > 400 nA.
Replace sensor.
yes
no
–
–
E 031
Failure
SCS fault reference
electrode 2
yes
no
–
–
E 032
Failure
Outside set slope
range for sensor 1
Sensor aged or defective;
Reference aged, defective or diaphragm
blocked;
Buffer solutions too old or contaminated;
PML not in the buffer solutions
yes
no
–
–
E 033
Failure
Outside set zero point
for sensor 1
yes
no
–
–
E 034
Failure
Outside set offset
range for sensor 1
yes
no
–
–
E 035
Failure
Outside set slope
range for sensor 2
Sensor aged or defective;
Reference aged, defective or diaphragm
blocked;
Buffer solutions too old or contaminated;
PML not in the buffer solutions
yes
no
–
–
E 036
Failure
Outside set zero point
for sensor 2
yes
no
–
–
E 037
Failure
Outside set offset
range for sensor 2
yes
no
–
–
E038
Mainte-
nance
Delta limit exceeded
Difference between channel 1 and 2
measured values too high. Process to
inconsistant or sensor defective. Replace
sensor if necessary.
yes
no
–
–
E040
Mainte-
nance
SCC / electrode sta-
tus of sensor 1 bad
Check sensor, replace if necessary;
maybe clean, (glass membrane blocked
or run dry; diaphragm blocked).
yes
no
–
E041
Mainte-
nance
SCC / electrode sta-
tus of sensor 2 bad
yes
no
–
E043
Mainte-
nance
Buffer difference
channel 1 too small
Wrong buffer used;
Buffer entry incorrect; buffer self-recogni-
tion defective.
yes
no
–
E044
Mainte-
nance
Meas. value channel
1 unstable
PAL missing; sensor too old; sensor some-
times dry; cable or plug defective.
yes
no
–
E045
Failure
Calibration aborted
Repeat calibration and renew buffer solu-
tion. Replace electrode if necessary.
yes
no
–
E048
Mainte-
nance
Buffer difference
channel 2 too small
Wrong buffer used;
Buffer entry incorrect; buffer self-recogni-
tion defective.
yes
no
–
E049
Mainte-
nance
Meas. value channel
2 unstable
PAL missing; sensor too old; sensor some-
times dry; cable or plug defective.
yes
no
–
E054
Mainte-
nance
Dosage time alarm
Dosage time exceeded at total dosage.
Dosage interrupted, dosing agent empty
or process too inconsistant.
yes
no
no
Error
no.
NAMUR
class
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning
start
Fact
User
Fact
User
Fact
User