5 Operation
Mycom S CPM 153
22
Hauser
5.1.4
Data logger
In the CPM 153, you have two data loggers available. With these data loggers you can
•
Record a parameter with 500 sequential measuring points or
•
two parameters each with 250 sequential measuring points.
To be able to use the function, activate the data logger in the "PARAM" menu
➞
"Set up 2"
➞
"Data Logger" (s. page 54). The function is active immediately.
You can view the measured values by scrolling through the different measuring menus
(see above).
– The current measured values are recorded in Record mode.
– In the "PARAM" menu
➞
"Set up 2"
➞
"Data logger" you can open saved data by
stating the date and time.
5.1.5
Operation access authorisation
To protect the transmitter against an unintended or undesired change in the
configuration and calibration data, functions can be protected using four-digit access
codes. Access authorisation has the following levels:
Display level (accessible without a code):
The complete menu can be viewed. The configuration cannot be altered. No calibration
is possible. On this level, only the controller parameters for new processes can be chan-
ged in the "DIAG" menu branch.
Maintenance code
Maintenance level (can be protected by the maintenance code):
This code allows access to the calibration menu.
Use this code to operate the temperature compensation item. The test functions and the
internal data can be viewed.
Factory setting Code = 0000, i.e. the levels are not protected.
You can ask for a universally valid maintenance code by contacting your service centre.
Specialist code
Specialist level (can be protected by the specialist code):
All menus can be accessed and changed.
Factory setting: Code = 0000, i.e. the levels are not protected.
You can ask for a universally valid specialist code by contacting your service centre.
To activate the codes (= functions locked) see the item "PARAM"
➞
"Set up 1"
➞
"Access
codes" (s. page 34).
Enter your desired code here. If the code is activated, you can only
edit the protected areas with the rights mentioned above.
!
Note!
•
Note down the selected code as well as the universal code and keep it in a place
where unauthorised persons do not have access to it.
•
If you set the code back to "0000", all the levels are freely accessible for editing.
The code can only be reset via the "specialist" menu.
Record mode
Scroll mode
2.00
pH1
12.00
Measure
Select
2.00
pH1
7.54
12 15 35
01 04 2001
12.00
Measure
Select