Mycom S CPM 153
6 Start-up
Hauser
37
– Linear delta pH on current output
If the current outputs are defined via positive and negative delta values, there is a
linear output on the current outputs (see table column 2).
•
Note on table entry: Enter the number of support points (value pairs) in Field EAB1.
Enter the value pairs in Field EAB2. For example. (4 support points):
ƒig. 17:
Example of characteristic enter using a table
Delta pH as value on current output
Linear delta pH on current output
mA
Delta pH
mA
Delta pH
0/4 mA
20 mA
0 pH
4 pH
0/4 mA
20 mA
–4 mA
4 pH
4
-4
mA
C07-CPM153xx-05-06-00-xx-009.EPS
20
pH
4
-4
mA
pH
C07-CPM153xx-05-06-00-xx-010.EPS
4
20
8
4
2
6 8
12
20
mA
16
pH
C07-CPM153xx-05-06-00-xx-002.EPS