Mycom S CPM 153
8 Troubleshooting
Hauser
93
8.1.1
Error number list: Trouble-shooting and configuration
In the following error list, you can find a description of all the error numbers occurring.
For each error number, you can see whether the error triggers
•
an alarm,
•
an error current or
•
cleaning
in the factory setting (=Fact.).
To enter the error list, proceed as follows:
!
Note!
•
Please process the errors as shown in Field H5 (Alarm menu) on page 44.
•
The second column shows the assignment acc. to NAMUR work sheet NA64 (failure,
maintenance, function control).
⇒
⇒
Diag
Edit
Next
Select
E
Hold
Error list
Error log
Operating log
Calibration log
Service
pH 7.00
Error
no.
NAMUR
class
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning
start
Fact
User
Fact
User
Fact
User
E001
Failure
Memory defective
Switch instrument off and on.
If necessary corrective maintenance at
factory.
yes
no
–
–
E002
Failure
Data error in EEP-
ROM
yes
no
–
–
E 003
Failure
Invalid configuration
Repeat download.
E004
Failure
Invalid hardware
code
The new software cannot recognise the
module.
E006
Failure
Transmitter 1
defective
test with new transmitter
yes
no
–
–
E007
Failure
Transmitter 2
defective
yes
no
–
–
E 008
Failure
SCS message
sensor 1
Impedance of pH glass membrane too
low: check pH sensor; replace it, if neces-
sary
For IsFET sensor: leak current > 400 nA.
Replace sensor.
yes
no
no
E 009
Failure
SCS message
sensor 2
yes
no
no
E 010
Failure
Temperature sensor 1
defective
Check temperature sensor and connec-
tions.
yes
no
no
E 011
Failure
Temperature sensor 2
defective
Check temperature sensor and connec-
tions.
yes
no
no
E019
Failure
Delta limit exceeded
Difference between channel 1 and 2
measured values too high. Process to
inconsistant or sensor defective. Replace
sensor if necessary.
yes
no
–
–