To enter the menu, proceed as follows:
⇒
⇒
PARAM
pH
7.00
Param
Parametrieren
Grundeinstellung
Sonderfunktionen
Handbedienung
Quick
Setup
Edit
Weiter E
pH
7.00
Hold
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next
E
pH
7.00
Param
Edit
Next
Set up 1
E
Hold
Relays
Temperature
Alarm
Hold
Calibration
CODE
CHOICE
(default = bold)
INFO
G1
Temperature
Medium compensation
Selection for temperature compensation
Temperature = automatic (ATC) or manual (MTC)
temperature compensation.
Medium compensation (only for pH) = compensation of
the medium using customer-specific tables (see below).
Temperature:
GA1
Measuring circuit 1
Measuring circuit 2
Select measuring circuit
you wish to configure.
Measuring circuit 1 (or 2, optional):
GAA1
ATC K1
ATC K2
MTC
MTC+Temp.
Select temperature compensation
ATC = automatic temperature compensation with a temp.
sensor circuit 1 or circuit 2
MTC = manual temp. comp. (with fixed temperature, ente-
red in Field GAA2)
MTC+Temp. = as MTC. On the display however, the value
which appears is what the temperature sensor measures
in the medium.
GAA2
025.0°C
(0 ... 100.0 °C)
MTC temperature
(only pH, MTC)
Temperature entry for manual compensation
GAA3
Off
On
Select temperature measurement
(only redox)
The reference temperature can be adapted according to
customer specifics in Field GBC1.
GAA4
Pt 100
Pt 1000
NTC 30k
Select temperature sensor
GAA5
Temperature actual
value
(–20.0 ... 150.0°C)
Enter actual value temperature for temperature
calibration
The value current measured by the temperature sensor
can be changed/adapted. The temperature difference is
stored internally as an offset value.
GAA6
0.0°C
(-5.0 ... 5.0°C)
Enter offset value
The offset value obtained from the previous field can be
edited or reset here.