6 Start-up
Mycom S CPM 153
68
Hauser
To enter the menu, proceed as follows:
6.5.14
Set up 2 - Controller quick adjustment
In this menu you can adjust the controller setpoint.
To enter the menu, proceed as follows:
⇒
⇒
PARAM
pH
7.00
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next
E
Hold
pH
7.00
Param
Edit
Next
Set up 2
E
Hold
Check systems
Data log
Contr. quick adj.
Limit switch
Controller settings
CODE
CHOICE
(default = bold)
INFO
P1
Limit switch 1
Limit switch 2
Limit switch 3
Limit switch 4
Limit switch 5
Selection
of the limit switch which you wish to configure. There are
five limit switches available.
Limit switch 1 / 2 / 3 / 4 / 5:
PA1 / PB1 /
PC1 / PD1 /
PE1
Function
Assignment
On point:
Off point:
off
pH/Redox
pH 16.00
(1500 mV/
100%/150 °C)
pH 16.00
(1500 mV/
100%/150 °C)
Limit switch configuration:
Function
: activation of function as limit switch
Assignment
: Selection of the measured value which shall
valid for the limit value. Selection: pH/redox, tempera-
ture, Delta (only if operating mode = redundancy)
On point
: Entry of the value at which the limit value
function is activated.
Off point
: Entry of the value at which the limit value
function is deactivated.
(Settable range: pH –2.00 ... 16.00 /
–1500 mV ... +1500 mV / 0 ... 100% / –50 ... +150 °C)
PA2 / PB2 /
PC2 / PD2 /
PE2
On delay:
Off delay:
Alarm
threshold:
0000 s
0000 s
pH 16.00
(150°C)
Limit switch configuration:
On delay
: Entry of the switch-on delay
(Range 0 ... 2000 s)
Off delay
: Entry of the switch-off delay
(Range 0 ... 2000 s)
Alarm threshold
: Entry of the value (alarm threshold) at
which the fault signalling contact switches.
⇒
⇒
⇒
PARAM
pH
7.00
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next
E
Hold
pH
7.00
Param
Edit
Next
Set up 2
E
Hold
Check systems
Controller settings
Limit switch
Data log
Contr. quick adj.
pH
7.00
Param
Select
Next
Quick adjustm
E
Hold
Function
set
7.00 pH
act.
6.58 pH