8 Troubleshooting
Mycom S CPM 153
96
Hauser
E055
Failure
Display range of main
parameter 1 under-
shot
Measuring line broken,
sensor in air or air cushion in assembly,
Potential matching missing in symmetrical
measurement, static charging in media
with lowest conductivity
yes
no
no
E056
Failure
Display range of main
parameter 2 under-
shot
yes
no
no
E057
Failure
Display range of main
parameter 1
exceeded
yes
no
no
E058
Failure
Display range of main
parameter 2
exceeded
yes
no
no
E059
Failure
Temperature range 1
undershot
Temperature sensor defective;
Sensor line interrupted or short-circuited;
Wrong sensor type selected
yes
no
no
E060
Failure
Temperature range 2
undershot
yes
no
no
E061
Failure
Temperature range 1
exceeded
yes
no
no
E062
Failure
Temperature range 2
exceeded
yes
no
no
E063
Mainte-
nance
Current limit
0/4 mA output 1
Measured value outside specified current
range:
Check measured value for plausibility,
if necessary adjust current output assign-
ment 0/4 mA and/or 20 mA.
yes
no
no
E064
Mainte-
nance
Current limit 20 mA
output 1
yes
no
no
E065
Mainte-
nance
Current limit 0/4 mA
output 2
yes
no
no
E066
Mainte-
nance
Current limit 20 mA
output 2
yes
no
no
E067
Mainte-
nance
Reference value
exceeded controller
LS 1
Dosing devices defective;
Chemical supply empty;
Measured value incorrect -> check for
plausibility and function;
Incorrect control direction set;
incorrect contact assigned;
Incorrect control function assigned
yes
no
no
E068
Mainte-
nance
Reference value
exceeded controller
LS 2
yes
no
no
E069
Mainte-
nance
Reference value
exceeded controller
LS 3
yes
no
no
E070
Mainte-
nance
Reference value
exceeded controller
LS 4
yes
no
no
E071
Mainte-
nance
Reference value
exceeded controller
LS 5
yes
no
no
Error
no.
NAMUR
class
Error message
Possible causes / measures
Alarm
contact
Error
current
Automatic
cleaning
start
Fact
User
Fact
User
Fact
User