Mycom S CPM 153
6 Start-up
Hauser
87
Calibration
redox relative
"Calibration relative"
For calibration, two tanks are filled with a sample of the medium. The contents of the first
tank are detoxified and are called Buffer 1.
The contents of the second tank are left unchanged and are called Buffer 2.
CODE
CHOICE
(default = bold)
INFO
C1
Electrode 1
Electrode 2
shared
Abort calibration
Selection for calibration
(only two-circuit)
Select electrode 1
or
2, and then run through calibra-
tion for each individual electrode.
CC
Calibration with
abs. calibration
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
CCD1 / CCD4
Immerse:
Electrode in buffer 1
Handling instructions
Immerse the electrode in buffer 1 / 2 (detoxified
sample, see above).
Confirm by pressing
F
CCD2 / CCD5
20 %
(0 ... 30 %)
Enter buffer
Enter the relative redox value of buffer 1 / 2 (detoxified
sample) in percent.
CCD3 / CCD6
Time: 10 s
mV 1:
225
Checks the stability of the calibration
Wait until the measurement is stable:
Time does not count down,
mV value no longer flashes,
Display "Measured value stable"
Confirm by pressing
F
CCD7
Invalid calibration value
Note in display:
Error message if the entered offset is too large.
CCD8
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing
F
accepts the new calibration data.
Reject
: The data is not accepted and the device has
not been recalibrated.
Recalibrate
: The data is rejected and the device is
recalibrated.
CCD9
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that measure-
ment can take place?