Mycom S CPM 153
6 Start-up
Hauser
29
T19
pH/redox K1
pH/redox K2
Temperature K1
Temperature K2
Select current output 1
(K2 only for two-circuit)
Selection of the parameter, which shall be
output on the current output.
T20
pH/Redox K1
pH/Redox K2
Temperature K1
Temperature K2
Delta
Continous controller
Select current output 2
(K2 and Delta only for two-circuit)
Selection of the parameter, which should
be output at the current output.
Delta:
The difference between the two
measuring circuits will be output at the cur-
rent output (circuit 1 – circuit 2).
Continous controller
: Control of a
controlling actuator via the current output
(See also Controller menu page 56).
T22
(0...9; A...Z)
Enter your customer specific instrument
number.
32-digit tag number.
This is saved in the DAT module which is
obtainable as an option.
T23
restart
end
Exit Quick Setup?
restart
= Run through settings in Fields T1-
T22 again
end
= Save the settings in Fields
T1-T22 and exit Quick Setup.
CODE
DISPLAY
CHOICE
(default = bold)
INFO
Hold
pH
7.00
Param
Edit
E
Output 1
pH/mV Input 1
pH/
2
mV Input
Next
Temperature Input1
Temperatur
2
e Input
Delta
Hold
pH
7.00
Param
Edit
E
Output 2
pH/mV Input 1
pH/
2
mV Input
Temperature Input1
Temperature Input2
Delta
Next
pH
7.00
Param
Edit
Hold
E
0...9, A...z
tag number
Next
Hold
pH
7.00
Param
Edit
E
Next
Start up
end
restart