Mycom S CPM 153
6 Start-up
Hauser
27
T6
Dual channel
Redundancy
Look-ahead
Selection (only two circuit)
Dual channel:
2 electrodes work comple-
tely independently of each other.
Redundancy:
Detection of electrode wear.
Look-ahead:
Early reaction to flow and pH
changes.
For description, see note s. page 30.
T7
Glass el. 7.0
Glass el. 4.6
Antimony
IsFET
Select electrode type 1
(only pH)´
!
Note!
• In the event of a change from glass
or antimony electrode to IsFET, the
temperature sensor is reset to Pt 1000 as
a default. Conversely, Pt 100 is selected.
• Dependent on the ordered measuring
input, "Glass el. 7.0" or "IsFET" is dis-
played as default value.
T8
Glass el. 7.0
Glass el. 4,6
Antimony
IsFET
Select electrode type 2
(only with pH, two circuit)
T9
solution ground
no solution ground
Select connection type
solution ground
= with potential matching
connection (PML)
no solution ground
= without PML
T10
°C
°F
Select temperature display
T11
ATC K1
ATC K2
MTC
MTC+Temp
Select temperature compensation K1
ATC = automatic temperature
compensation
MTC = manual temp. comp. (with fixed
temperature, entered in Field GAA2)
MTC+Temp. = as MTC. On the display
however, the value which appears is what
the temperature sensor measures in the
medium.
T12
025.0°C
Temperature value K1
(only with pH and selection of MTC or
MTC+Temp. in Field T11)
CODE
DISPLAY
CHOICE
(default = bold)
INFO
Hold
Edit
EE
pH
7.00
Param
Dual 1+2
Dual channel
Redundancy
Look-ahead
Next
Hold
pH
7.00
Param
Edit
E
Antimon
IsFET
pH electr.typeK1
Next
Glass El. 7.0
Glass El. 4.6
IsFET
Hold
pH
7.00
Param
Antimon
Edit
E
pH electr.typeK2
Next
Glass El. 7.0
Glass El. 4.6
Hold
E
Edit
pH
7.00
Param
Sensor ground
solution ground
no solution ground
Hold
pH
7.00
Param
Edit
E
°C
°F
Temp. unit
Hold
pH
7.00
Param
Edit
E
ATC K2
MTC
MTC+Temp
ATC K1
Temp.comp. K1
Next
Hold
pH
7.00
Param
Edit
-20.0...150.0°C
E
0
25.0
MTC-Temp. K1
Next