6 Start-up
Mycom S CPM 153
76
Hauser
Service data / logbooks:
YAA1
0000
Service code entry required
!
Note!
For service access code, see Field D1, p. 34.
YAA2
Note in display:
Incorrect service code entry
(back to the last field)
Simulations:
YB1
Simulation:
Output:1
Output:2
off
12.00 mA
00.00 mA
Adapt simulation (current outputs)
Simulation off
: The frozen values from the last measure-
ment are used for the simulation.
Simulation on
: The current values for the outputs can
be changed (Output 1, Output 2)
YB2
Simulation:
Measured
value 1:
Temperature:
Measured
value 2:
Temperature:
off
pH 07.00
025.00°C
pH 00.00
000.00°C
Adapt simulation (measured value/temperature)
Simulation off
: The frozen values from the last measure-
ment are used for the simulation.
Simulation on
: The values (measured value/tempera-
ture) can be changed.
YB3
Simulation:
Failure contact:
Contact 1:
Contact 2:
...
off
off
off
off
Adapt simulation (contacts)
Simulation off
: The last statuses are frozen and used for
the simulation.
Simulation on
: The contacts can either be opened (on)
or closed (off).
!
Note!
If you return to the measurement mode with the simula-
tion switched on, "Simul" and "Hold" flash in the display.
Check systems
YC1
Display
Keypad
RAM
EEPROM
Flash
Select check
Display
: All the fields are queries alternately. Defective
cells become visible.
Keypad
: All the keys must be pressed one after the
other. If the system is functioning perfectly, the approp-
riate symbols appear in the display.
RAM
: "RAM O.K" message if there are no errors.
EEPROM
: "EEPROM O.K" message if there are no
errors.
Flash
(memory): "Flash OK" message if there are no
errors.
!
Note!
Leave this item by pressing "PARAM".
CODE
CHOICE
(Default = bold)
INFO