6 Start-up
Mycom S CPM 153
50
Hauser
JC8
threshold
length
02 mV
010s
Stability
During calibration the mV values may maximally vary for
the given threshold within the defined time range (length).
During calibration, the mV value may change during the
given period ("duration") at maximum by the stated
amount ("threshold"), so that the calibration is
considered as stable.
Therefore you can adjust accuracy and timing
individually to your process.
Calibration timer:
JD1
Cal timer:
on
Warning:
0001h
___________
Time: 0001:00
Calibration timer
If no calibration is undertaken in the set time, an error
message appears (E115).
Cal Timer
: on = activate
Warning
: Enter the time within which a calibration must
take place.
Time
: Display of the remaining time to an error message.
CODE
CHOICE
(default = bold)
INFO