Mycom S CPM 153
6 Start-up
Hauser
83
Calibration
redox absolute
"Calibration absolute"
The transmitter has a calibrated mV display range.
One
absolute mV value is set with
a single buffer solution (adaptation of the measuring chain offset). A buffer solution
preferably with 225 or 475 mV is used.
CODE
CHOICE
(default = bold)
INFO
C1
Electrode 1
Electrode 2
shared
Abort calibration
Selection for calibration
(only two-circuit)
Select electrode 1
or
2, and then run through
calibration for each individual electrode.
CB
Calibration with
abs. calibration
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
CBB1
Immerse:
Electrode in buffer
Handling instructions
Immerse the electrode in the buffer.
Confirm by pressing
F
CBB2
0225 mV
(–1500 ... +1500 mV)
Enter buffer
During calibration, enter the buffer mV value.
CBB3
Time: 10 s
mV 1:
225
Checks the stability of the calibration
Wait until the measurement is stable:
Time does not count down,
mV value no longer flashes,
Display "Measured value stable"
Confirm by pressing
F
CBB4
Invalid calibration value
Note in display:
Error message if the entered offset is too large.
CBB5 / CBB6
Offset
Good
0005
mV
Note in display:
Info on electrode 1.
Data on the offset and calibration quality.
CBB7
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing
F
accepts the new calibration data.
Reject
: The data is not accepted and the device has
not been recalibrated.
Recalibrate
: The data is rejected and the device is
recalibrated.
CBB8
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that measure-
ment can take place?