5.6
Probe passes calibration, but still appears to
read too high
Potential cause
There may be a leak that is permitting ambient air to mix with the process gases. Since
many combustion processes are slightly negative in pressure, ambient air caqn be sucked
into the cell area, biasing the O
2
reading upward.
Recommended actions
1. Ensure that the calibration gas line is capped tightly between calibrations. If using
Autocal, make sure to check the valve seating properly.
2. If an abrasive shield is installed to protect the entire probe from particulate erosion,
a leak in the probe flange gasket can allow ambient air to migrate down the annular
space between the probe and shield and then into the cell. Always install a new
probe flange gasket when reinstalling a probe.
3. Check calibration gas hoses or tubing for leaks. Repair leaks in hose or tubing.
4. Check cell flange corrugated seal for leaks. Replace seal.
Potential cause
There may be a leak inside the probe itself, permitting the reference air (20.95 percent O
2
)
to mix with the process gases at the cell.
Recommended actions
1. To confirm this leak condition, connect instrument air for reference. Pressurize the
inside (reference side) of the probe by plugging the reference air exhaust port with
your finger for one minute.
The O
2
reading should decrease slightly. If the O
2
reading increases during this test,
there is a leak inside the probe.
Acid condensation inside the probe can degrade the red silcone tube that carries
the cal gas to the cell.
) to inspect this hose.
See
.
Troubleshoot
Reference Manual
October 2019
00809-0100-4880
70
Emerson.com/Rosemount
Summary of Contents for Rosemount Oxymitter DR
Page 4: ...4 ...
Page 52: ...Startup and operate Reference Manual October 2019 00809 0100 4880 52 Emerson com Rosemount ...
Page 72: ...Troubleshoot Reference Manual October 2019 00809 0100 4880 72 Emerson com Rosemount ...
Page 74: ...Return material Reference Manual October 2019 00809 0100 4880 74 Emerson com Rosemount ...
Page 84: ...Replacement parts Reference Manual October 2019 00809 0100 4880 84 Emerson com Rosemount ...