4. Replace with new EPROM.
5. Close electronics enclosure and power up system.
2.4.5
Set up World Class 3000 intelligent field transmitter
The Rosemount
™
Oxymitter DR probe operates with a 115 Vac heater. Ensure that the
voltage selection jumpers in the IFT or HPS, if used, are set properly.
IFT 3000 power supply board jumper configuration
WARNING
Always disconnect line voltage from intelligent field transmitter before changing jumpers.
Line voltage selection
Jumper (install)
100 Vac
JM3, JM7, JM2
120 Vac
JM8, JM7, JM1
220 Vac
JM6, JM5, JM2
240 Vac
JM6, JM5, JM1
Probe heater voltage selection
Jumper (install)
World Class probe (44 V)
JM10
218 probe (115 V)
JM9
World Class Direct Replacement probe (115 V)
or Rosemount Oxymitter Direct Replacement
probe (115 V)
JM9
CAUTION
If incorrect heater voltage is selected, damage to the probe may occur.
Always update the relevant labeling to reflect the select voltage.
Heater power supply (HPS 3000) jumper configuration
WARNING
Always disconnect line voltage from heater power supply and analog electronics (if used)
before changing jumpers.
Line voltage selection
Jumper (install)
100/120 Vac
JM4, JM1
220/240 Vac
JM5
(1)
100 Vac operation requires transformer part number 1M03961G02.
Reference Manual
Install
00809-0100-4880
October 2019
Rosemount Oxymitter DR
47
Summary of Contents for Rosemount Oxymitter DR
Page 4: ...4 ...
Page 52: ...Startup and operate Reference Manual October 2019 00809 0100 4880 52 Emerson com Rosemount ...
Page 72: ...Troubleshoot Reference Manual October 2019 00809 0100 4880 72 Emerson com Rosemount ...
Page 74: ...Return material Reference Manual October 2019 00809 0100 4880 74 Emerson com Rosemount ...
Page 84: ...Replacement parts Reference Manual October 2019 00809 0100 4880 84 Emerson com Rosemount ...