4
Maintenance and service
4.1
Overview
This section identifies the calibration methods available and provides the procedures to
maintain and service the transmitter.
WARNING
When working on the laboratory bench, be aware that the transmitter, probe tube, and
flame arrestor hub can be hot (up to 572 °F [300 °C]) in the region of the probe heater.
WARNING
Failure to install covers and ground leads could result in serious injury or death.
Install all protective equipment covers and safety ground leads after installation.
4.2
Calibrate
Calibrate the Rosemount
™
Oxymitter DR when you commission it.
WARNING
The probe may be hot and may cause severe burns.
Remove the probe from the stack for all service activities. Allow the unit to cool and take it
to a clean work area.
WARNING
There is voltage up to 115 Vac.
Disconnect and lock out power before working on any electrical components.
Under normal circumstances, the probe will not require frequent calibration. When
calibration is required, follow the procedure in the Reference Manual applicable to your
electronics package. Use
to track probe performance.
Reference Manual
Maintenance and service
00809-0100-4880
October 2019
Rosemount Oxymitter DR
53
Summary of Contents for Rosemount Oxymitter DR
Page 4: ...4 ...
Page 52: ...Startup and operate Reference Manual October 2019 00809 0100 4880 52 Emerson com Rosemount ...
Page 72: ...Troubleshoot Reference Manual October 2019 00809 0100 4880 72 Emerson com Rosemount ...
Page 74: ...Return material Reference Manual October 2019 00809 0100 4880 74 Emerson com Rosemount ...
Page 84: ...Replacement parts Reference Manual October 2019 00809 0100 4880 84 Emerson com Rosemount ...