Oxymitter oxygen transmitters (Integral & DR): OXT4C, OXT4CNF, OXT4CDR, and
OXT4CDRNF
Class I, Division 1, Groups B, C, and D
Class I, Zone 1, Ex d IIB+H2 T2
Class I, Zone 1, AEx d IIB+H2 T2
Oxymitter oxygen transmitters (split achitecture, probe with remote electronics): OXT4C,
OXT4CNF, 6A00094G
Class I, Zone 1, Ex de IIB+H2 T6 (remote electronics) / T2 probe)
Class I, Zone 1, AEx de IIB+H2 T6 (remote electronics) / (T2 (probe)
Class 1, Division 1, Groups B, C, and D T2 (probe)
Class 1, Division 2, Groups B, C and D T6 (remote electronics)
Special Conditions for Safe Use:
1. Non-flame arrestor probe -"NF" versions must have the probe installed in a non-
classified area.
2. When the probe tube is mounted in a hazardous (classified) area (with flame
arrestor end to complete the assembly) calibration lines that travel in and out of the
equipment bringing reference gas must not contain a pressure higher than 1.1
times the atmospheric pressure; these calibration lines shall not contain pure
oxygen, acetylene, or combustible gases other than the gases for which this Zone 1
classified application has been investigated for: Group IIB +H2 gases.
3. There are no interchangeable components.
4. Contact the original manufacturer for information of flameproof joint dimensions.
5. Fastener property class must be A2-70 for units intended for use in Zone 1 classified
areas.
6. Any other electronics not covered by this report and used in conjunction with the
Rosemount
™
Oxymitter probes or with the remote electronics must be installed in a
non-classified area (ex. Rosemount Xi Advanced Electronics).
(1)
Except Temperature Code T1 for 44V
∼
model of Rosemount OXT4CDR only.
Rosemount
™
Oxymitter
™
product certifications
Reference Manual
October 2019
00809-0100-4880
92
Emerson.com/Rosemount
Summary of Contents for Rosemount Oxymitter DR
Page 4: ...4 ...
Page 52: ...Startup and operate Reference Manual October 2019 00809 0100 4880 52 Emerson com Rosemount ...
Page 72: ...Troubleshoot Reference Manual October 2019 00809 0100 4880 72 Emerson com Rosemount ...
Page 74: ...Return material Reference Manual October 2019 00809 0100 4880 74 Emerson com Rosemount ...
Page 84: ...Replacement parts Reference Manual October 2019 00809 0100 4880 84 Emerson com Rosemount ...