a) Connect the oxygen signal lines from the signal conditioning electronics to
b) Connect the Type K thermocouple lines from the signal conditioning
electronics to the terminals indicated in
4. Install terminal cover (20,
) and secure with cover lock (27), gasket (26),
and screw (25).
2.3
Pneumatic installation
If instrument air will be used as reference air, connect the reference air set to the
transmitter.
CAUTION
Damage can result from having a cold transmitter exposed to the process gases.
Upon completing installation, make sure the transmitter is turned on and operating before
firing up the combustion process.
CAUTION
Thermal cycling can cause condensation and premature aging.
During outages, if possible, leave all transmitter units running.
CAUTION
If washing down the ducts during outage, make sure to power down the transmitter units
and remove them from the wash area.
. Install the reference air set in accordance with
Install
Reference Manual
October 2019
00809-0100-4880
40
Emerson.com/Rosemount