Table 2-8: Replacement Parts
(continued)
Part name
Pressure
Part number
3
Combination filter-
regulator
0-30 psig
4505C21G01
Instrument air (reference air): 10 psig (68.95 kPa) minimum, 225 psig (1551.38 kPa)
maximum at 2 scfh (56.6 L/hour) maximum; less than 40 parts per million total
hydrocarbons. Set regulator outlet pressure at 5 psa (35 kPa).
2.4
Set up probe with electronics package
You can use the Rosemount
™
Oxymitter DR probe with several electronics packages,
including: Models 218, 218A, 225, 132 (analog and digital), TC200, and ZA8C.
WARNING
Failure to install covers and ground leads could result in serious injury or death.
Install all protective equipment covers and safety ground leads before starting the
equipment.
2.4.1
Set up Westinghouse Models 218, 225, and 132
(analog) electronics
Change the probe heater setpoint of the existing electronics to support the Rosemount
™
Oxymitter DR probe.
Procedure
1. Open electronics enclosure.
2. On temperature controller card, connect jumper wire from TP3 to either Pin 2 or Pin
7.
Install
Reference Manual
October 2019
00809-0100-4880
42
Emerson.com/Rosemount