Figure 2-15: Air Set, Plant Air Connection
Schematic hookup for reference air supply on Rosemount
™
Oxymitter DR probe head.
Dimensions are in inches with millimeters in parentheses.
A. 0.125-27 NPT female outlet connection
B. Outlet
C. 0.25-18 NPT female inlet connection
D. Two mounting holes 3.19 (81.03) LG through body for 0.312 (7.92) diameter bolts
E. Instrument air supply: 10-225 psig maximum pressure
F. Reference air set: 263C152G01
G. To electronics
H. 0.250 or 6 mm O.D. tubing (supplied by customer)
I. 0.250 or 6 mm O.D. tube compression fitting
J. Drain valve
K. Flow set point knob
Table 2-8: Replacement Parts
Part name
Pressure
Part number
1
Flowmeter
0.2-2.0 scfh
771B635H02
2
2-in. (50.8 mm)
pressure gauge
0-15 psig
275431-006
Reference Manual
Install
00809-0100-4880
October 2019
Rosemount Oxymitter DR
41