Relative humidity
5 to 95% (non-condensing)
Signals
Analog output/HART: 4-20 mA isolated from power supply, 950
ohms maximum load
Logic I/O: Two terminal logic contact configurable as either an
alarm output or as a bi-directional calibration handshake signal
to Rosemount IMPS 4000 or SPS 4001B, self-powered (+5V) in
series with 340 ohms conduit ports -
¾
-in. -14 NPT (for analog
output and logic I/O signal lines)
Power requirements
Normal operation: 175 W
Maximum (warmup): 500 W
Ambient operating
temperature (junction box)
200 °F (93 °C)
160 °F (71 °C) maximum for YEW replacement
Rosemount has satisfied all obligations coming from the
European legislation to harmonize the product requirements in
Europe.
1.4.2
Physical specifications
Materials of
construction
Transmitter probe: Process-wetted materials are 316L or 304 stainless
steel.
Rosemount
™
Oxymitter Transmitter electronics housing (integral to
probe or remote mounted): Low copper aluminum IP66 (NEMA 4X)
with reference air exhaust port piped to clean area
Optional Rosemount Xi Electronics: NEMA 4X, polycarbonite material
Process
connections
2-in. 150# (4.75-in. [121 mm] bolt circle); DIN (5.71-in. [145 mm] bolt
circle)
Note
Flanges are flat-faced and for mounting only. Flanges are not
pressure-rated. Spool piece part numbers are available to offset
electronics housing from hot ductwork. Many adapter flanges are
available to mate to existing flanges.
Orientation
Vertical or horizontal mount
Table 1-7: Probe Lengths and Approximate Shipping Weights
Length
Weight
18 in. (457 mm)
16 lb. (7.3 kg)
3 ft. (0.91 m)
21 lb. (9.5 kg)
6 ft. (1.83 m)
27 lb. (12.2 kg)
9 ft. (2.74 m)
33 lb. (15.0 kg)
12 ft. (3.66 m)
39 lb. (17.7 kg)
15 ft. (4.57 m)
45 lb. (20.5 kg)
18 ft. (5.49 m)
51 lb. (23 kg)
Description and specifications
Reference Manual
October 2019
00809-0100-4880
22
Emerson.com/Rosemount
Summary of Contents for Rosemount Oxymitter DR
Page 4: ...4 ...
Page 52: ...Startup and operate Reference Manual October 2019 00809 0100 4880 52 Emerson com Rosemount ...
Page 72: ...Troubleshoot Reference Manual October 2019 00809 0100 4880 72 Emerson com Rosemount ...
Page 74: ...Return material Reference Manual October 2019 00809 0100 4880 74 Emerson com Rosemount ...
Page 84: ...Replacement parts Reference Manual October 2019 00809 0100 4880 84 Emerson com Rosemount ...