6. Ensure the transmitter is properly earthed by way of both internal and external
points.
CAUTION
Uninsulated stacks or ducts may cause ambient temperatures around the
electronics to exceed 185 °F (85 °C), which may cause overheating damage to the
electronics.
If you remove duct work insulation for transmitter mounting, replace the insulation
afterwards. See
. Ensure the probe installation does not obsure the
warnings on the housing covers.
2.2
Electrical installation with integral electronics
All wiring must conform to local and national codes.
WARNING
Disconnect and lock out power before connecting the unit to the power supply.
WARNING
Failure to install covers and ground leads could result in serious injury or death.
Install all protective equipment covers and safety ground leads after installation.
WARNING
To meet the safety requirements of IEC 61010 (EC requirement) and ensure safe
operation of this equipment, connect to the main power supply through a circuit
breaker (minimum 10 A) which will disconnect all current-carrying conductors during a
fault situation.
Make sure this circuit breaker includes a mechanically operated isolating switch.
If there is no switch, locate another external means of disconnecting the power supply
close by.
Circuit breakers or switches must comply with a recognized standard, such as IEC
60947.
WARNING
The instrument is heavy.
Use proper lifting and carrying procedures to avoid personal injury.
Install
Reference Manual
October 2019
00809-0100-4880
36
Emerson.com/Rosemount
Summary of Contents for Rosemount Oxymitter DR
Page 4: ...4 ...
Page 52: ...Startup and operate Reference Manual October 2019 00809 0100 4880 52 Emerson com Rosemount ...
Page 72: ...Troubleshoot Reference Manual October 2019 00809 0100 4880 72 Emerson com Rosemount ...
Page 74: ...Return material Reference Manual October 2019 00809 0100 4880 74 Emerson com Rosemount ...
Page 84: ...Replacement parts Reference Manual October 2019 00809 0100 4880 84 Emerson com Rosemount ...