CAM216/16 IOM
Doc. Number: ECM-402-0116
January 2016, Rev. 1
37
Section 6: Device Variables
Device Variables
Section 6: Device Variables
6.1
Device Variables
This Field Device has nine (9) Device Variables. These variables can be selected against the
HART primary variables: PV, SV, TV, and QV with some limitations.
Table 25.
Device Variables Table
Device Variable Code
Device Variable Name
Units
0
Setpoint
%
1
Position
%
2
Torque
%
3
Analog Input 1
%
4
Analog Input 2
%
5
Analog Output 1
%
6
Analog Output 2
%
7
Hydraulic Pressure
PSI (Pounds Per Square Inch)
8
Accumulator Pressure
PSI (Pounds Per Square Inch)
6.2
Dynamic Variables
Table 26.
Dynamic Variables Table
Sr. No.
Dynamic Variable
Default Assignment Valid Range
1
Primary Variable
0: Setpoint
0: Setpoint
2
Secondary Variable
1: Position
0: Setpoint
1: Position
3
Tertiary Variable
2: Torque
2: Torque
3: Analog Input 1
4: Analog Input 2
5: Analog Output 1
6: Analog Output 2
7: Hydraulic Pressure
8: Accumulator Pressure
4
Quaternary Variable
3: Analog Input 1
2: Torque
3: Analog Input 1
4: Analog Input 2
5: Analog Output 1
6: Analog Output 2
7: Hydraulic Pressure
8: Accumulator Pressure