CAM216/16 IOM
Doc. Number: ECM-402-0116
January 2016, Rev. 1
7
Section 2: Installation of CAM16
Installation of CAM16
2.
Lower the TBM 320A board without disconnecting the wires attached on the sides and
bottom. This will provide access to the 320B board.
Figure 3
Accessing the 320B Board
3.
Locate the J2 connector on the 320B Board, and connect to the J3 connection on the
CAM16. Align the J2 to J3 and the 3 white plastic holders to mount the CAM16 in place.
4.
Follow Step 5, 6 or 7 depending on desired HART Network type:
Table 6.
HART Control Types (1)
Control Type
Step to Follow
HART – Analog Control (Point-to-Point)
Step 5
HART – Digital Control (Multi-Drop)
Step 6
HART – Digital Control (Wireless THUM)
Step 7
5.
HART Analog Control (Point-to-Point; Actuator Must be in Analog Control Mode)
a.
Connect the HART Analog 4 – 20 mA input signal as follows:
Table 7.
HART Control Types (2)
4 – 20 mA
Connector J4, PIN 3
4 – 20 mA
Connector J4, PIN 4