Figure 3-9: Transmitter electronic enclosure security latch
A. Transmitter Electronics Enclosure end cap
B. Security latch
Procedure
1. Rotate the end cap clockwise fully closing and compressing the end cap seal. Install
the Security latch using a 3mm Allen wrench.
2. Install the security seal wire into and through one of the two holes in the end cap.
Choose holes that minimize counterclockwise rotation of the end cap when the
security wire is taut (maximum wire diameter .078 inch; 2.0mm).
Figure 3-10: Transmitter Electronics Enclosure security seal installation
A. Transmitter Electronics Enclosure end cap
B. Security wire seals
3. Adjust the security wire, removing all slack and thread into the lead seal.
4. Cut wire ends to remove excess wire.
3.7.2
Base enclosure security seals
Use the following instructions to install the security seal wire on the Base Enclosure.
Installation manual
Electrical Installation
3-9000-760
May 2019