Figure 4-19: AMS Device Manager - Service Tools Trends
Primary and Secondary variables display real-time uncorrected volume flow
rate trends. The third and fourth variables charts display trends for
temperature and pressure.
15. Click the
Service Tools
→
Routine Maintenance
tab. Click
Analog Output 1 Trim
to perform a digital to analog trim adjustment of the first milliampere output. The
4mA and 20mA output current values should equal the plant’s standard values.
a) Click
Yes
to confirm the configuration changes. Repeat this step to trim
Analog Output 2 current.
b) Click
Apply
to write the output trim values to the meter.
c) Click
OK
to navigate back to the Service Tools page.
16. Click the
Service Tools
→
Zero Calibration
tab. See AMS Device Manager - Guided
to configure the zero flow parameters.
17. After you have changed and written the configuration changes to the meter do the
following:
a) Enable the Write Protect switch on the CPU Module to protect the meter’s
configuration.
b) Replace the end cap and if required, apply security seals through the endcap
holes and through the hex head bolts that secure the Bracket/Cover to the
Base enclosure.
Note
The next time you connect to the device using Daniel MeterLink, the Monitor
page displays a Meter status alarm that the configuration has changed and
Configuration
Installation manual
May 2019
3-9000-760
88
3814 Liquid Ultrasonic Flow Meter
Summary of Contents for Daniel 3814
Page 1: ...Installation manual 3 9000 760 Rev HA May 2019 Daniel 3814 Liquid Ultrasonic Flow Meter ...
Page 2: ...2 ...
Page 10: ...Contents Installation manual May 2019 3 9000 760 x 3814 Liquid Ultrasonic Flow Meter ...
Page 92: ...Configuration Installation manual May 2019 3 9000 760 92 3814 Liquid Ultrasonic Flow Meter ...
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......