for an external power source. Refer to the Field wiring diagram DMC-004936 in
.
3.6.3
Analog output settings
The 960-24
™
MSTS provides two 4-20 mA analog output signals that are software
configurable for either sink or source current (see
DMC-004936).
Full HART
®
functionality is provided so that any commercially available HART
®
transmitter
which meets the specifications of the HART
®
Communications Foundation can be
connected to the Daniel 3814 Liquid Ultrasonic Flow Meter.
Analog Output 2 (AO2) is user-configurable as a conventional 4-20 mA output.
3.6.4
Digital input
The 960-24
™
MSTS provides one digital input that can be used as a general purpose input
or used for synchronizing calibration (for flow calibration gating-contact closure). The
meter records the volume seen between switch closures. The polarity of the input is
configured as normal or inverted polarity.
•
Polarity is determined by the
IsDI1ForCalActiveLow
and the gating edge is determined
by the
IsDI1ForCalStateGated
(calibrate edge gated or calibrate state gated).
•
Calibration is started via an inactive>active state change.
The digital input must be configured via the Daniel MeterLink
Tools
→
Edit/Compare
Configuration
page.
3.6.5
DHCP server switch settings
The meter can be configured to act as a DHCP server. The DHCP server is enabled/disabled
via CPU Module DHCP switch as follows:
Table 3-6: DHCP server switch settings
CPU Module switch
DHCP server disabled
DHCP server enabled
DHCP
OFF
ON
3.6.6
Configuration protect switch settings
The meter’s configuration parameters and firmware can be protected against changes via
CPU Module Write PROT. switch as follows:
Table 3-7: Configuration protect switch settings
CPU Module switch
Configuration protected
Configuration unprotected
WRITE PROT.
ON (default)
OFF
A complete list of write-protected parameters is in
Installation manual
Electrical Installation
3-9000-760
May 2019
Installation manual
63
Summary of Contents for Daniel 3814
Page 1: ...Installation manual 3 9000 760 Rev HA May 2019 Daniel 3814 Liquid Ultrasonic Flow Meter ...
Page 2: ...2 ...
Page 10: ...Contents Installation manual May 2019 3 9000 760 x 3814 Liquid Ultrasonic Flow Meter ...
Page 92: ...Configuration Installation manual May 2019 3 9000 760 92 3814 Liquid Ultrasonic Flow Meter ...
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......