installed in the meter body. The process temperature must also not exceed the operating
temperature range of the transducers. LT-01 and LT-03 transducers have an operating
range from -50 °C (-58 °F) to +100 °C (+212 °F).
For process fluid temperatures up to 150 °C, the high temperature extended length
transducer cables (P/N 1-360-01-800) shall be used to connect the Daniel 3814 Liquid
Ultrasonic Flow Meter electronics to the transducers installed in the meter body. The
process temperature for LT-04, LT-05, LT-08 and LT-09 transducers must also not exceed
the operating temperature range of the transducers -50 °C (-58 °F) to +150 °C (+302 °F).
CAUTION
SURFACE TEMPERATURE HAZARD
The meter body and piping may be extremely hot or cold. Wear appropriate personal
protective equipment when coming in contact with the meter.
Failure to comply may result in injury.
Installation manual
Mechanical Installation
3-9000-760
May 2019
Installation manual
47