Table 1-2: Meter specifications
(continued)
Liquid Meter specifications
Accuracy Limits
•
Accuracy limits typically are:
—
±0.20% without a flow calibration
Power
Meter
•
10.4 VDC to 36 VDC
•
11 W typical power consumption
Temperature flameproof Transmitter Electronic
Enclosure and Base Electronics Enclosure
Serial cable
•
Belden #9940 or equivalent (22 gauge)
—
Capacitance (pF/m) 121.397 (conductor
to conductor)
—
Capacitance (pF/m) 219.827 (conductor
to other conductor and shield)
—
Resistance (DC) DCR @ 20 °C (Ohm/km)
48.2307 Nominal Outer shield resistance
- DCR @ 20 °C (Ohm/km) 16.405
—
Operating voltage - 300 V RMS (UL AWM
Style 2464)
—
Current 2.4 Amps per conductor @ 25 °C
(recommended)
Ethernet cable
•
Cat-5 Standard 100Mbps
)
•
22 AWG wire characteristics are as follows:
—
Capacitance = 20 pF/ft or 20 nF/1000 ft
(between two wires)
—
Resistance = 0.0168 Ohms/ft or 16.8
Ohms/1000 ft
—
Pull-up voltage is 24 VDC
Installation manual
Introduction
3-9000-760
May 2019
Installation manual
21
Summary of Contents for Daniel 3814
Page 1: ...Installation manual 3 9000 760 Rev HA May 2019 Daniel 3814 Liquid Ultrasonic Flow Meter ...
Page 2: ...2 ...
Page 10: ...Contents Installation manual May 2019 3 9000 760 x 3814 Liquid Ultrasonic Flow Meter ...
Page 92: ...Configuration Installation manual May 2019 3 9000 760 92 3814 Liquid Ultrasonic Flow Meter ...
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......