Table 3-2: CPU Module labeling and LED functions
(continued)
CPU Module labeling and LED
functions
Function
Switch position indicator or
LED
PORT A
•
PORT A override - RS-232
serves as an override during
meter commissioning to
establish communications
and in the event the user
cannot communicate with
the meter due to an
inadvertent
communication
configuration change. The
override period is for two
minutes
•
Supports:
—
auto-detected ASCII
(Start bit 1, Data Bit 7,
Parity Odd/Even, Stop
Bit 1)
—
RTU (Start Bit 1, Data Bit
8, Parity none, Stop Bit
1).
—
Modbus protocols
•
RS-232 Baud rate=19,200
•
Modbus ID=32
Switch position
•
ON - enables RS-232 PORT
A override
•
OFF - (default setting)
disables RS-232 PORT A
MEAS
System color indicates
metrology mode
•
Acquisition mode
•
Measurement mode
LED status
•
Red flashing LED
•
Solid red the Acquisition
Module not communicating
with the CPU Module
•
Green flashing LED
PWR
•
3.3V Power Indicator
•
Solid Green
LED 4
•
Not used
LED 5
•
Not used
RX
•
RX signal (Port A for RS485
or RS232 communication)
receiving data
•
Flashing green (when
receiving data)
TX
•
TX signal (Port A for RS485
or RS232 communication)
transmitting data
•
Flashing green (when
transmitting data)
LINK
•
ETH1Link user Ethernet
connection
•
Solid green
Installation manual
Electrical Installation
3-9000-760
May 2019
Installation manual
55
Summary of Contents for Daniel 3814
Page 1: ...Installation manual 3 9000 760 Rev HA May 2019 Daniel 3814 Liquid Ultrasonic Flow Meter ...
Page 2: ...2 ...
Page 10: ...Contents Installation manual May 2019 3 9000 760 x 3814 Liquid Ultrasonic Flow Meter ...
Page 92: ...Configuration Installation manual May 2019 3 9000 760 92 3814 Liquid Ultrasonic Flow Meter ...
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......