Figure 3-6: CPU Module - Frequency/Digital inputs common ground - Type 2
A. FODO1 and Digital input1 - shared common ground (Group 1)
B. FODO2 and FODO3 - shared common ground (Group 2)
Figure 3-7: CPU Module - Frequency/Digital outputs common ground - Type 4
A. FODO 1 and DI1/FODO6 - shared common ground - Type 4 CPU Module (Group1)
B. FODO2, FODO3, FODO4 and FODO5 - shared common ground - Type 4 CPU Module
(Group2)
3.6.2
Analog input settings
The The 960-24
™
MSTS has the capability to sample analog temperature (Analog Input 1)
and pressure (Analog Input 2) with 4-20 mA signals. These analog input signals are
configured to sink. The two independent analog input circuits are configured for
conventional 4-20 mA service. Also, 24VDC isolated power supply connection is provided
Electrical Installation
Installation manual
May 2019
3-9000-760
62
3814 Liquid Ultrasonic Flow Meter
Summary of Contents for Daniel 3814
Page 1: ...Installation manual 3 9000 760 Rev HA May 2019 Daniel 3814 Liquid Ultrasonic Flow Meter ...
Page 2: ...2 ...
Page 10: ...Contents Installation manual May 2019 3 9000 760 x 3814 Liquid Ultrasonic Flow Meter ...
Page 92: ...Configuration Installation manual May 2019 3 9000 760 92 3814 Liquid Ultrasonic Flow Meter ...
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......