11. Test parameter
Full-electronic HV-Tester F1-1x Typs
Max current rise delta I (A/ms)
: - enter the max. rate of change of cur-
rent allowed. The limit is:
0-50 mA/ms range 1 mA
0-500 mA/ms range 10 mA
0-5000 mA/ms range =>100 mA
Contact monitoring „on/off
“ - switches on the evaluation for the minimum
current. During the test a pre-defined minimum current must flow. This
ensures contact is made to the test object.
off
- no evaluation of the minimum current.
on
- minimum current monitoring active.
Minimum current (mA)
: - the current entered must flow during the HV
testing as a minimum. If the current is less than this value, a contact
fault (I<Imin) is indicated.